Fault | possible reason | Solution |
(1) Changes in retention time | 1 Column temperature change | Column constant temperature, if necessary, a constant temperature box should be configured |
2 Insufficient balance between isocratic and gradient | Equilibrate the column with at least 10 column volumes of mobile phase | |
3 The buffer capacity is not enough | Use >25mmol/L buffer | |
4 Column Contamination | Rinse column daily | |
5 Changes in column conditions | Stable injection conditions, adjust mobile phase | |
6 column nearing end of life | guard column | |
(2) The retention time is shortened | 1 flow rate increase | Check pump, reset flow rate |
2 Sample overload | Reduce sample volume | |
3 Loss of bonded phase | Keep the pH value of the mobile phase at 3~7.5 and check the direction of the column | |
4 Changes in mobile phase composition | Prevent mobile phase evaporation or precipitation | |
5 temperature increase | Column thermostat | |
(3) Extended retention time | 1 flow rate drop | The pipeline leaks, replace the pump sealing ring, and remove the air bubbles in the pump |
2. Changes of active sites on silica gel column | Use a mobile phase modifier, such as adding triethylamine, or use a base to passivate the column | |
3 Loss of bonded phase | Same as before (2) 3 | |
4 Changes in mobile phase composition | Same as before (2) 4 | |
5 temperature reduction | Same as before (2) 5 | |
(4) Shoulder peaks or bifurcations appear | 1 The sample volume is too large | Sample preparation with mobile phase, the total sample volume is less than 15% of the first peak |
2 The sample solvent is too strong | Use weaker sample solvents | |
3 The column collapses or forms a short-circuit channel | Replace column, use less aggressive conditions | |
4 Failure of sintered stainless steel in column | Replace the sintered stainless steel, add an in-line filter, and filter the sample | |
5 Injector damaged | Replace the Injector Rotor | |
(5) Ghost Peak | 1. Injection valve residual peak | Rinse the valve with a strong solvent after each use for improved valve and sample cleaning |
2. Unknown substances in the sample | Processing samples | |
3. The column is not equilibrated | Re-equilibrate the column and use the mobile phase as the sample solvent (especially for ion-pair chromatography) | |
4. Trifluoroacetic acid (TFA) oxidation (peptide mapping) | Freshly formulated every day, with antioxidants | |
5. Water pollution (reverse phase) | Check water quality by varying equilibration time, use HPLC grade water | |
(6) Baseline noise | 1 bubble (sharp peak) | Degas the mobile phase, add column back pressure |
2 Pollution (random noise) | Wash column, clean up sample, use HPLC grade reagents | |
3 Detector lamp continuous noise | Replacing the Deuterium Lamp | |
4 Electrical interference (occasional noise) | Use regulated power supply, check the source of interference (such as water bath, etc.) | |
5 Bubbles in the detector | Degas the mobile phase, add column back pressure | |
(7) Peak tailing | 1 column overload | Reduce sample volume, increase column diameter and use higher capacity stationary phase |
2 peak interference | Clean samples, adjust mobile phase | |
3 The role of silanol | Add triethylamine, use alkali-induced passivation column to increase the concentration of buffer or salt, reduce the pH value of mobile phase, and passivate the sample | |
4 Same as before (4) 4 | Same as before (4) 4 | |
5 Same as before (4) 3 | Same as before (four) 3 | |
6 Dead volume or extra column volume is too large | Keep connection points to a minimum, make appropriate adjustments for all connection points, and use as small an internal diameter as possible for connection tubing | |
7 Decreased column efficiency | With less corrosive conditions, replace the column and use a guard column | |
(8) Peak broadening | 1 Injection volume is too large | Same as (4) 1 |
2 Causes peak extension in the injection valve | Air bubbles are removed before and after injection to reduce dispersion | |
3 The sampling rate of the data system is too slow | Set the rate so that each peak is greater than 10 points | |
4 The detector time constant is too large | Set the time constant to 10% of the half-width of the first peak of interest | |
5 mobile phase viscosity is too high | Increase column temperature, use low viscosity mobile phase | |
6 The volume of the detection cell is too large | With a small volume pool, remove the heat exchanger | |
7 retention time is too long | Gradient elution can also be used with increasing solvent content during isocratic elution | |
8 Extra column volume too large | Minimize connecting pipe diameter and connecting pipe length | |
9 Sample overload | Inject small concentration and small volume samples |