Determination and Solution of Common Faults in High Performance Liquid Chromatography

2022-10-19 11:53

Fault

possible reason

Solution

(1) Changes in retention time

1 Column temperature change

Column constant temperature, if necessary, a constant temperature box should be configured

2 Insufficient balance between isocratic and gradient

Equilibrate the column with at least 10 column volumes of mobile phase

3 The buffer capacity is not enough

Use >25mmol/L buffer

4 Column Contamination

Rinse column daily

5 Changes in column conditions

Stable injection conditions, adjust mobile phase

6 column nearing end of life

guard column

(2) The retention time is shortened

1 flow rate increase

Check pump, reset flow rate

2 Sample overload

Reduce sample volume

3 Loss of bonded phase

Keep the pH value of the mobile phase at 3~7.5 and check the direction of the column

4 Changes in mobile phase composition

Prevent mobile phase evaporation or precipitation

5 temperature increase

Column thermostat

(3) Extended retention time

1 flow rate drop

The pipeline leaks, replace the pump sealing ring, and remove the air bubbles in the pump

2. Changes of active sites on silica gel column

Use a mobile phase modifier, such as adding triethylamine, or use a base to passivate the column

3 Loss of bonded phase

Same as before (2) 3

4 Changes in mobile phase composition

Same as before (2) 4

5 temperature reduction

Same as before (2) 5

(4) Shoulder peaks or bifurcations appear

1 The sample volume is too large

Sample preparation with mobile phase, the total sample volume is less than 15% of the first peak

2 The sample solvent is too strong

Use weaker sample solvents

3 The column collapses or forms a short-circuit channel

Replace column, use less aggressive conditions

4 Failure of sintered stainless steel in column

Replace the sintered stainless steel, add an in-line filter, and filter the sample

5 Injector damaged

Replace the Injector Rotor

(5) Ghost Peak

1. Injection valve residual peak

Rinse the valve with a strong solvent after each use for improved valve and sample cleaning

2. Unknown substances in the sample

Processing samples

3. The column is not equilibrated

Re-equilibrate the column and use the mobile phase as the sample solvent (especially for ion-pair chromatography)

4. Trifluoroacetic acid (TFA) oxidation (peptide mapping)

Freshly formulated every day, with antioxidants

5. Water pollution (reverse phase)

Check water quality by varying equilibration time, use HPLC grade water

(6) Baseline noise

1 bubble (sharp peak)

Degas the mobile phase, add column back pressure

2 Pollution (random noise)

Wash column, clean up sample, use HPLC grade reagents

3 Detector lamp continuous noise

Replacing the Deuterium Lamp

4 Electrical interference (occasional noise)

Use regulated power supply, check the source of interference (such as water bath, etc.)

5 Bubbles in the detector

Degas the mobile phase, add column back pressure

(7) Peak tailing

1 column overload

Reduce sample volume, increase column diameter and use higher capacity stationary phase

2 peak interference

Clean samples, adjust mobile phase

3 The role of silanol

Add triethylamine, use alkali-induced passivation column to increase the concentration of buffer or salt, reduce the pH value of mobile phase, and passivate the sample

4 Same as before (4) 4

Same as before (4) 4

5 Same as before (4) 3

Same as before (four) 3

6 Dead volume or extra column volume is too large

Keep connection points to a minimum, make appropriate adjustments for all connection points, and use as small an internal diameter as possible for connection tubing

7 Decreased column efficiency

With less corrosive conditions, replace the column and use a guard column

(8) Peak broadening

1 Injection volume is too large

Same as (4) 1

2 Causes peak extension in the injection valve

Air bubbles are removed before and after injection to reduce dispersion

3 The sampling rate of the data system is too slow

Set the rate so that each peak is greater than 10 points

4 The detector time constant is too large

Set the time constant to 10% of the half-width of the first peak of interest

5 mobile phase viscosity is too high

Increase column temperature, use low viscosity mobile phase

6 The volume of the detection cell is too large

With a small volume pool, remove the heat exchanger

7 retention time is too long

Gradient elution can also be used with increasing solvent content during isocratic elution

8 Extra column volume too large

Minimize connecting pipe diameter and connecting pipe length

9 Sample overload

Inject small concentration and small volume samples