Phenomenon | inspection plan | Check phenomenon/result | possible reason | Solution | |
1 No sample peak appears | Check if the current is stable | no current | Capillary clogged or broken | Rinse the capillary with water and observe whether there is water flowing out. If there is no water flowing out, please remove the cartridge and check whether the ends of the capillary and the window are broken; The buffer solution needs to be filtered, and the sample is filtered or centrifuged to remove particles. | |
The current fluctuates wildly until it almost disappears | There are bubbles in the buffer solution or sample precipitation in the zone | Ultrasonic degas the buffer solution. If this phenomenon still occurs, it may be that there is precipitation in the sample area. You can try to solve this problem by reducing the sample concentration/prolonging the ramp time; for samples with low solubility in the buffer solution, Additives need to be added to the buffer solution to solve this problem. | |||
The initial value of the current is small, and then gradually increases | Sample injection volume is too large | Reduce the injection volume, usually the injection parameters are set at about 0.5psi, 5sec | |||
normal current | Sample concentration too low | Use high-concentration sample test, if it cannot be solved, it may be the following other reasons | |||
The detection wavelength is not set correctly | Please confirm the characteristic absorption of the analyte and check the detection wavelength setting in the method | ||||
separation polarity error | For protein samples, please pay attention to the PI and charge of the protein under separation conditions; for nucleic acid samples, it will be negatively charged under normal conditions | ||||
The sample is adsorbed on the inner wall of the capillary | For protein and nucleic acid samples, try to use coated capillary tubes for separation, or use extreme pH conditions or dynamic coatings to prevent sample adsorption. | ||||
Damaged optical detector or fiber | Carry out the test of the standard sample, if there is no corresponding result, there may be a hardware problem, please contact the engineer | ||||
Check capillary window | Whether there is a transparent window | Forgot to open the capillary window or the window position is not correct | Re-open the capillary detection window, or adjust the window to the correct position | ||
2 Tailing of sample peaks | The sample is adsorbed on the inner wall of the capillary | For protein and nucleic acid samples, try to use coated capillary tubes for separation, or use extreme pH conditions or dynamic coatings to prevent sample adsorption. | |||
3 The peak shape of the sample is asymmetric | capillary inlet | Capillary entry cut not flush | Re-cut the capillary inlet, pay attention to the capillary cutting method, do not use too much force or repeatedly scratch | ||
Change sample solvent | Buffer solution and sample solution The difference in electrolytic rate is too large | Buffer solution can be used as sample solvent | |||
4 Sample peak is too broad | Reduce sample injection volume | has improved | Sample concentration is too high | Can reduce sample concentration or reduce injection volume | |
no improvement | The nature of the sample itself is not uniform | This reason is mainly for protein samples, and the probability of small molecule samples is low | |||
5 Migration time is unstable | Peak time is unstable and irregular | The buffer solution equilibrates slowly with the inner wall of the capillary | Avoid flushing the capillary with NaOH between each sample run | ||
Peak departure times are delayed in turn | There are substances in the sample that are prone to adsorption | First rinse the capillary with 0.1N NaOH for a short time before each sample run to see if the migration time is repeatable, if the effect is not good, use a coated capillary to improve the results or reprocess the sample to remove easily adsorbed components in the sample | |||
6 Peak shape and peak time are unstable | Change the type of buffer solution | Peak shape and time stability | The buffer solution is unstable, easy to electrolyze, or the sample is unstable under the original buffer solution conditions | Change to other types of buffer solutions | |
Peak shape and peak extraction are still unstable | Substances in the sample are easily adsorbed on the inner wall of the capillary | Coated capillaries can be used to overcome this problem, or to prepare the sample | |||
7 Current Leakage | Check if the capillary is broken at the window | capillary break | The outflow of the buffer solution in the capillary causes the current leakage | Replace the capillary with a new one, and clean the fiber optic head and detection window with water | |
Capillary without break | Excessive humidity in the experimental environment | Use a dehumidifier and open the Cartridge Cover of the instrument; if there is no dehumidifier, you can use the air conditioner, and if the humidity is too high, you can place the desiccant when the instrument is turned off, but be sure to take out the desiccant when the instrument is running | |||
8 Unable to add air pressure | Check if the cork is tight or replace the cork | Problem solved after replacement | Cork aging, loss of elasticity | Please buy new corks and note that the corks are not tumble dry, only dry naturally | |
There is liquid in the neck of the bottle, causing the cork to slip easily | Do not overfill the bottle with liquid and do not wet the neck of the bottle | ||||
If there is no problem with the cork, the vacuum method can be used first, if the vacuum method can be used, but the pressure is still not available | pressure system problems | Please contact an engineer | |||
9 The migration time is reproducible but the peak area reproducibility is not good | Recut capillary ends | If the problem is solved | uneven capillary entrance | Re-cut the capillary inlet, pay attention to the capillary cutting method, do not use too much force or repeatedly scratch | |
Problem persists, try motorized injection | If motorized injection remains reproducible, there is a problem with the pressure control or pneumatics | Please contact an engineer | |||
10 Capillary or electrode is easy to break | Check whether the bottle cap is aging | bottle cap aging | buy new bottle caps | ||
Whether the electrodes are not vertical | Electrodes are not vertical | Straighten the electrodes | |||
11 Coolant leaks | Check the gasket and clips inside the cartridge | Missing or wrong order | Install in strict accordance with the installation sequence in the operation manual | ||
The internal pipeline of the instrument is not tightly sealed | Please contact an engineer | ||||
12 PDA low light intensity | Detection window Aperture model | Aperture with narrow slits | Please be sure to use the Aperture marked 8 when using the DAD detector | ||
13 The spectral baseline is unstable | Rinse the capillary with 0.1N NaOH before running the original method without injection | baseline improvement | The substance in the sample is adsorbed | Re-pretreat samples or change electrophoresis conditions | |
Baseline not improving | There is strong adsorption in the capillary, or the buffer system is unstable | Replace with a new capillary, do not inject, only run the buffer, observe the baseline | |||
If the baseline is still unstable after replacing a new capillary, it can be tested with Run Buffer A | baseline improvement | The buffer solution equilibrates slowly with the inner wall of the capillary | Change the type of buffer solution, if the baseline change does not interfere with the detection, the original electrophoresis conditions can be maintained | ||
No improvement from baseline | Detect irregularities in light sources or other optics | Please contact an engineer | |||