FAQ of Capillary Electrophoresis Experiment

2022-10-19 11:55

Phenomenon

inspection plan

Check phenomenon/result

possible reason

Solution

1 No sample peak appears

Check if the current is stable

no current

Capillary clogged or broken

Rinse the capillary with water and observe whether there is water flowing out. If there is no water flowing out, please remove the cartridge and check whether the ends of the capillary and the window are broken; The buffer solution needs to be filtered, and the sample is filtered or centrifuged to remove particles.

The current fluctuates wildly until it almost disappears

There are bubbles in the buffer solution or sample precipitation in the zone

Ultrasonic degas the buffer solution. If this phenomenon still occurs, it may be that there is precipitation in the sample area. You can try to solve this problem by reducing the sample concentration/prolonging the ramp time; for samples with low solubility in the buffer solution, Additives need to be added to the buffer solution to solve this problem.

The initial value of the current is small, and then gradually increases

Sample injection volume is too large

Reduce the injection volume, usually the injection parameters are set at about 0.5psi, 5sec

normal current

Sample concentration too low

Use high-concentration sample test, if it cannot be solved, it may be the following other reasons

The detection wavelength is not set correctly

Please confirm the characteristic absorption of the analyte and check the detection wavelength setting in the method

separation polarity error

For protein samples, please pay attention to the PI and charge of the protein under separation conditions; for nucleic acid samples, it will be negatively charged under normal conditions

The sample is adsorbed on the inner wall of the capillary

For protein and nucleic acid samples, try to use coated capillary tubes for separation, or use extreme pH conditions or dynamic coatings to prevent sample adsorption.

Damaged optical detector or fiber

Carry out the test of the standard sample, if there is no corresponding result, there may be a hardware problem, please contact the engineer

Check capillary window

Whether there is a transparent window

Forgot to open the capillary window or the window position is not correct

Re-open the capillary detection window, or adjust the window to the correct position

2 Tailing of sample peaks



The sample is adsorbed on the inner wall of the capillary

For protein and nucleic acid samples, try to use coated capillary tubes for separation, or use extreme pH conditions or dynamic coatings to prevent sample adsorption.

3 The peak shape of the sample is asymmetric

capillary inlet


Capillary entry cut not flush

Re-cut the capillary inlet, pay attention to the capillary cutting method, do not use too much force or repeatedly scratch

Change sample solvent


Buffer solution and sample solution

The difference in electrolytic rate is too large

Buffer solution can be used as sample solvent

4 Sample peak is too broad

Reduce sample injection volume

has improved

Sample concentration is too high

Can reduce sample concentration or reduce injection volume

no improvement

The nature of the sample itself is not uniform

This reason is mainly for protein samples, and the probability of small molecule samples is low

5 Migration time is unstable


Peak time is unstable and irregular

The buffer solution equilibrates slowly with the inner wall of the capillary

Avoid flushing the capillary with NaOH between each sample run

Peak departure times are delayed in turn

There are substances in the sample that are prone to adsorption

First rinse the capillary with 0.1N NaOH for a short time before each sample run to see if the migration time is repeatable, if the effect is not good, use a coated capillary to improve the results or reprocess the sample to remove easily adsorbed components in the sample

6 Peak shape and peak time are unstable

Change the type of buffer solution

Peak shape and time stability

The buffer solution is unstable, easy to electrolyze, or the sample is unstable under the original buffer solution conditions

Change to other types of buffer solutions

Peak shape and peak extraction are still unstable

Substances in the sample are easily adsorbed on the inner wall of the capillary

Coated capillaries can be used to overcome this problem, or to prepare the sample

7 Current Leakage

Check if the capillary is broken at the window

capillary break

The outflow of the buffer solution in the capillary causes the current leakage

Replace the capillary with a new one, and clean the fiber optic head and detection window with water

Capillary without break

Excessive humidity in the experimental environment

Use a dehumidifier and open the Cartridge Cover of the instrument; if there is no dehumidifier, you can use the air conditioner, and if the humidity is too high, you can place the desiccant when the instrument is turned off, but be sure to take out the desiccant when the instrument is running

8 Unable to add air pressure

Check if the cork is tight or replace the cork

Problem solved after replacement

Cork aging, loss of elasticity

Please buy new corks and note that the corks are not tumble dry, only dry naturally

There is liquid in the neck of the bottle, causing the cork to slip easily

Do not overfill the bottle with liquid and do not wet the neck of the bottle

If there is no problem with the cork, the vacuum method can be used first, if the vacuum method can be used, but the pressure is still not available

pressure system problems

Please contact an engineer

9 The migration time is reproducible but the peak area reproducibility is not good

Recut capillary ends

If the problem is solved

uneven capillary entrance

Re-cut the capillary inlet, pay attention to the capillary cutting method, do not use too much force or repeatedly scratch

Problem persists, try motorized injection

If motorized injection remains reproducible, there is a problem with the pressure control or pneumatics

Please contact an engineer

10 Capillary or electrode is easy to break

Check whether the bottle cap is aging


bottle cap aging

buy new bottle caps

Whether the electrodes are not vertical


Electrodes are not vertical

Straighten the electrodes

11 Coolant leaks

Check the gasket and clips inside the cartridge


Missing or wrong order

Install in strict accordance with the installation sequence in the operation manual

The internal pipeline of the instrument is not tightly sealed

Please contact an engineer

12 PDA low light intensity

Detection window Aperture model


Aperture with narrow slits

Please be sure to use the Aperture marked 8 when using the DAD detector

13 The spectral baseline is unstable

Rinse the capillary with 0.1N NaOH before running the original method without injection

baseline improvement

The substance in the sample is adsorbed

Re-pretreat samples or change electrophoresis conditions

Baseline not improving

There is strong adsorption in the capillary, or the buffer system is unstable

Replace with a new capillary, do not inject, only run the buffer, observe the baseline

If the baseline is still unstable after replacing a new capillary, it can be tested with Run Buffer A

baseline improvement

The buffer solution equilibrates slowly with the inner wall of the capillary

Change the type of buffer solution, if the baseline change does not interfere with the detection, the original electrophoresis conditions can be maintained

No improvement from baseline

Detect irregularities in light sources or other optics

Please contact an engineer