Capillary electrophoresis (CE), also known as high performance capillary electrophoresis (HPCE), is one of the fastest developing analytical methods in recent years. In 1981, Jorgenson and Lukacs first proposed that μ Modern capillary electrophoresis (CE) was established by using high voltage for separation in m inner diameter capillary column. Terabe et al. established micellar capillary electrokinetic chromatography in 1984. In 1987, Hjerten established capillary isoelectric focusing, and Cohen and Karger proposed capillary gel electrophoresis. From 1988 to 1989, the first commercial capillary electrophoresis instruments appeared. In just a few years, CE has been developing rapidly because it meets the separation and analysis requirements for polypeptides, proteins (including enzymes and antibodies), nucleotides and even deoxyribonucleic acid (DNA) in various fields of life science represented by bioengineering.
CE is a combination of classical electrophoresis technology and modern microcolumn separation. Compared with high performance liquid chromatography (HPLC), CE is similar in that both are high performance separation technologies, instrument operation can be automated, and both have many different separation modes. The difference between the two is that CE replaces the retention time in HPLC with migration time, and the analysis time of CE usually does not exceed 30 min, which is faster than HPLC; For CE, it is theoretically deduced that the theoretical tray height is proportional to the diffusion coefficient of solute. For biomacromolecules with small diffusion coefficient, the column efficiency is much higher than that of HPLC; The sample required for CE is nl grade, with a minimum of 270fl. The amount of mobile phase is only a few milliliters, while the sample required for HPLC is μ Level l, mobile phase requires hundreds of milliliters or more; However, CE can only be prepared in small quantities, while HPLC can be prepared in large quantities.
CE is much faster than ordinary electrophoresis because of its high electric field; In addition to failing to connect with atomic absorption spectroscopy and infrared spectroscopy, other types of detectors have been connected with CE for detection; Generally, the quantitative accuracy of electrophoresis is poor, while CE and HPLC are similar; CE operation is much more automatic than ordinary electrophoresis. In a word, the advantages of CE can be summarized as three high and two low: high sensitivity, detection limit of common ultraviolet detector can reach 10-13~10-15mol, and that of laser induced fluorescence detector can reach 10-19~10-21mol; High resolution, with hundreds of thousands of theoretical plates per meter; The higher one can reach millions or even tens of millions, while HPLC is generally thousands to tens of thousands; With high speed, it can be completed in 60s as soon as possible, 10 proteins can be separated in 250s, 19 cations can be separated in 1.7min, and 30 anions can be separated in 3min; There are few samples, and only nl (10-9 L) level of sample injection is required; Low cost, only a small amount (several milliliters) of mobile phase and cheap capillary. Because of the above advantages and the ability to separate biological macromolecules, CE has become one of the fastest developing separation and analysis methods in recent years. Of course, CE is still a developing technology, and some theoretical research and practical applications are under development.
"CE" generally refers to a kind of liquid phase separation technology driven by high voltage electric field and separated by capillary tube according to the difference in mobility and distribution behavior of each component in the sample. The instrument structure includes a high-voltage power supply, a capillary tube, a detector and two buffer solution storage bottles for inserting both ends of the capillary tube and connecting with the power supply. In electrolyte solution, the phenomenon that charged particles migrate in the opposite direction of their charges at different speeds under the action of electric field is called electrophoresis.
When pH>3, the inner surface of the quartz capillary column used for CE is negatively charged, forming a double electric layer when contacting with the solution.
Under the action of high voltage, the phenomenon that the water combined cation in the electric double layer causes the fluid to move towards the negative electrode as a whole is called electroosmosis. The migration speed of particles in the electrolyte in the capillary is equal to the vector sum of electrophoresis and electroosmosis (EOF). The movement direction of positive ions is consistent with that of electroosmosis, so they flow out first; The electrophoretic flow velocity of neutral particles is "zero", so their migration velocity is equivalent to that of electroosmotic flow; The movement direction of negative ions is opposite to that of electroosmotic flow. However, since the electroosmotic flow velocity is generally higher than that of electrophoretic flow, it will flow out after neutral particles, so that separation can be achieved due to different migration velocities of various particles.
Electroosmosis is the driving force to push the fluid forward in CE. It makes the whole fluid move forward at a uniform speed like a plug, making the whole flow pattern approximate to a flat "plug flow". It makes the solute zone not expand in the capillary in principle. However, in HPLC, the pressure driven method makes the fluid in the column parabolic, and the velocity at the center is twice the average velocity, leading to the expansion of the solute zone itself, resulting in a decrease in column efficiency, which makes the separation efficiency inferior to CE.
Theoretical analysis shows that increasing speed is an important way to reduce spectral band broadening and improve efficiency, and increasing electric field strength can increase speed. However, the high electric field strength causes the current to increase, which causes the electrolyte in the capillary to generate Joule heat (self heating). Self heating will cause the fluid to have a parabolic temperature distribution in the radial direction, that is, the temperature at the center of the pipe axis is higher than that near the wall. As the solution viscosity decreases exponentially with the increase of temperature, the temperature gradient makes the medium viscosity gradient in the radial direction, which affects the solute migration speed, making the solute molecules in the center of the tube axis migrate faster than those near the tube wall, resulting in the broadening of the spectral band and the decrease of the column efficiency.
Generally speaking, the mobility will increase by 2% when the temperature is increased by 1 ℃ (the so-called mobility refers to the distance of solute moving in unit time interval and unit electric field). In addition, the change of temperature changes the pH value and viscosity of the solution, further leading to changes in electroosmotic flow, charge distribution of solute molecules (including protein structure), ionic strength, etc., resulting in changes in mobility, poor repeatability, and decreased column efficiency. Decreasing the buffer concentration can reduce the current intensity and reduce the temperature difference. High ionic strength buffer solution can prevent protein from being adsorbed on the tube wall, and can produce concentration focusing effect on the column to prevent peak expansion and improve peak shape. Reducing the pipe diameter alleviates the heat accumulation caused by the high electric field to a certain extent, but the small pipe diameter reduces the injection volume, resulting in technical difficulties in injection and detection. Therefore, accelerating heat dissipation is an important way to reduce the temperature difference caused by self heating. The thermal conductivity of liquid is 100 times higher than that of air. At present, some capillary electrophoresis instruments with liquid cooling can use buffer solution with ionic strength up to 0.5mol/L for separation, or use 200 μ M diameter capillary can still achieve good separation effect and reproducibility.