High performance capillary electrophoresis (HPCE) is a new type of liquid phase analysis technology with capillary as the separation channel and high voltage DC electric field as the driving force. It is another major progress in analytical science after high performance liquid chromatography (HPLC). It enables analytical science to move from the microliter level to the nanoliter level, and makes single cell and even single molecule analysis possible. It also provides new ideas for the separation and analysis of biological macromolecules, such as proteins, that have puzzled us for a long time. The instrument can be used to analyze organic compounds, inorganic ions, neutral molecules, chiral compounds, proteins and peptides, DNA and nucleic acid fragments. It is widely used in environmental protection, biochemistry, clinical testing, justice, food hygiene, drug testing, agricultural chemistry and other fields.
Principle of Analysis
The principle of ultraviolet detection is based on the difference between the absorbance of the tested component and the background electrolyte. When the tested component passes through the detection window, the change of absorbance follows Lambert Beer's law, that is, under certain experimental conditions, the absorbance is proportional to the concentration of the tested component.
1、 Performance and technical requirements
detector
1. Optical path system: diffraction grating monochromator, photocell
2. Lamp source: imported deuterium lamp, edge filtering, 370nm (tungsten lamp optional)
3. Automatic switching system of deuterium lamp and tungsten lamp
4. Wavelength range: 190-740nm
5. Bandwidth: Δλ ≤8nm
6. Wavelength accuracy: ± 2nm
7. Sensitivity: 2 × 10-5AU at 254nm, t=1s
8. GLP function: detailed report, lighting time and lighting times
9. Noise: ± 1.5 × 10-5AU
10. Baseline drift: 4 × 10-4AU/h
11. Detection limit: ≤ 1 × 10-6g/ml (p-nitrophenol at 318nm)
12. Quantitative repeatability: RSD% ≤ 3.0%
13. Qualitative repeatability: RSD% ≤ 1.3%
14. Minimum capillary length: 25 cm
high-voltage power supply
1. The voltage is directly controlled by software
2. Output voltage range: 0-35KV (standard configuration)
3. Output current range: less than 0-1000 less than 0-1000 μ A μ A
4. Voltage output error: ≤ 0.5%
5. Current output error: ≤ 0.5%
6. Time drift: ≤ 0.05%
7. Temperature drift: ≤ 0.05%
8. One click high voltage and data acquisition
9. High separation efficiency, fast analysis speed and high detection sensitivity
10. Low reagent consumption and simple sample pretreatment
11. Wide detection range
12. Short circuit protection: closing section self recovery protection
13. Variable wavelength selection, 190-740nm continuously adjustable
14. Injection mode: electric injection 0-35KV adjustable,
15. LCD screen display, touch the panel
16. Wavelength scanning, automatic zeroing, easy to learn